Hospital Refrigerator Manufacturers introduces the precautions for the use of ultra-low temperature refrigerators:
 
Hospital Refrigerator Manufacturers introduces the precautions for the use of ultra-low temperature refrigerators:

1. Indoor temperature: 5-32℃, relative humidity 80%/22℃.
2. Distance from the ground > 10cm. The altitude is below 2000m.
3. It takes 6 hours to decrease from +20°C to 80°C.
4. Strong acid and corrosive samples should not be frozen.
5. Frequently check the sealing strip of the outer door.
6. Landing on all four feet is stable and level.
7. When there is a power failure prompt, press the stop beeping button.
8. The general refrigeration temperature is set at 60℃
9. The power supply voltage of 220v (AC) must be stable, and the power supply current must be at least 15A (AC) or more.
10. In the event of a power failure, the power switch and battery switch at the back of the refrigerator must be turned off. When the normal power supply is restored, the power switch at the back of the refrigerator must be turned on, and then the battery switch shall be turned on.
11. Note that heat dissipation is very important to the refrigerator. It is necessary to maintain indoor ventilation and a good heat dissipation environment, and the ambient temperature cannot exceed 30C.
12. In summer, adjust the set temperature to -70°C, and pay attention to the usual setting not to be too low.
13. Don't open the door too big when accessing the samples, and keep the access time as short as possible.
14. Note that the samples that are frequently accessed should be placed on the upper second layer, and the samples that need to be stored for a long time infrequently accessed should be placed on the lower second layer, so as to ensure that the air-conditioning is not excessively lost when the door is opened, and the temperature will not rise too fast.
15. Note that the filter must be cleaned once a month (first vacuum with a vacuum cleaner, rinse with water after vacuuming, and finally dry and reset). The internal condenser must be vacuumed with a vacuum cleaner every two months.
16. Do not use force to open the door when the door is locked to avoid the door lock from being damaged.
17. To defrost, only cut off the power of the refrigerator and open the door. When the ice and frost begin to melt, a clean and absorbent cloth must be placed on each layer of the refrigerator to absorb and wipe the water (note that there will be a lot of water).

Through the above introduction, China Protective Mask (FFP2 NR) Manufacturers hopes that you can simply refer to the content of this article in future use.

https://www.jingeao.com/
-- END ---
Share Facebook Twitter
Print Friendly and PDF DisclaimerReport Abuse
Contact Email [email protected]
Issued By jingeaoningbo
Country China
Categories Accounting , Affiliate Program , Architecture
Tags hospital refrigerator , protective mask ffp2 nr
Last Updated December 20, 2021